ViperWire Advance Peripheral Guide Wire with Flex Tip, Model No. VPR-GW-FT18. Sterilized with Ethylene Oxide. The guide wire is a smooth, stainless steel wire, with a silicone coating, and a radiopaque distal spring tip. The guide wire allows for proper positioning of the device crown within peripheral arteries and provides a center of rotation for the device drive shaft. The guide wire torquer is a small, plastic accessory, packaged with the guide wire, and provides a gripping surface for manipulating the guide wire, if desired.

This recall has been terminated (originally issued March 3, 2017).

Company
Cardiovascular Systems Inc
Recall Initiated
March 3, 2017
Terminated
November 28, 2017
Recall Number
Z-1771-2017
Quantity
186 devices
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

Cardiovascular Systems, Inc. is recalling one lot of ViperWire Advance with Flextip, model VPR-GW-FT18 because it may contain a 0.014" ViperWire Advance Flextip product instead of the 0.018" ViperWire Advance with Flextip.

Distribution

Nationwide distribution in the states of AR, AZ, CO, FL, GA, HI, IN, KS, KY, LA, MS. NC, NM, RI, TX, VA

Lot / Code Info

Lot 156032

Root Cause

Under Investigation by firm

Action Taken

Customers were sent an Urgent Medical Device Recall letter beginning March 03, 2017. The letter identified affected product, reason for recall, and asked for affected devices to be removed from use and return to CSI. Questions can be directed to CSI Customer Service at 877-274-0901, or your CSI Sales Representative.