MOSAIQ Oncology Information System

Company
Elekta, Inc.
Recall Initiated
July 15, 2016
Recall Number
Z-2293-2016
Quantity
383
Firm Location
Atlanta, GA

Reason for Recall

It is possible that a change to an Order Set will not be saved in the current open Care Plan. This can result in a wrong chemo dose in the order.

Distribution

Worldwide distribution. US nationwide, Antigua and Barbuda, Australia, Austria, Bahamas, Bahrain, Canada, Czech Republic, Germany, Hungary, South Africa, Spain, Sweden, Turkey, and United Kingdom

Lot / Code Info

Versions 2.50.05 and higher

Root Cause

Software Design Change

Action Taken

Elekta sent Important Field Safety Notice 371-03-MSQ-001 to all customers on 7/15/2016. The notice informs users of the specific product and version numbers affected by the issue, and any work around that can be used to avoid the issue. The letter also included a response form which is to be returned to Elekta.