MOSAIQ Oncology Information System

Company
Elekta, Inc.
Recall Initiated
June 17, 2016
Terminated
August 18, 2020
Recall Number
Z-2052-2016
Quantity
69
Firm Location
Atlanta, GA

Reason for Recall

It is possible that that the selected Plan-of-the Day could be accidently changed by scrolling the mouse wheel up and down and inadvertently selecting another plan.

Distribution

US Distribution including Puerto Rico and to the states of : AK, AZ, CA, FL, GA, IL, IN, LA, MD, MI, MO, NJ, NY, OH, OR, PA, TN, TX, UT, VA, WA, and to the countries of : Canada, Denmark, Finland, Germany, Israel, Japan, Netherlands, Norway, Sweden, United Kingdom

Lot / Code Info

Versions 2.62 and higher

Root Cause

Software design

Action Taken

Elekta sent an Important Field Safety Notice to all customers on June 17, 2016. The notice informs users of the specific product and version numbers affected by the issue, and any work around that can be used to avoid the issue. A software patch will be issued for Mosaiq version 2.62 in order to the correct the defect. For further questions please call (404) 993-5886.