Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices.

This recall has been terminated (originally issued July 4, 2018).

Company
Medtronic Inc.
Recall Initiated
July 4, 2018
Terminated
June 6, 2022
Recall Number
Z-1282-2019
Quantity
Approximately 9,328 registered users
Firm Location
Northridge, CA
Official Source
View on FDA website ↗

Reason for Recall

The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.

Distribution

Distribution was only made OUS. Countries involved are: Austria, United Arab Emirate, Australia, Belgium, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Iceland, Italy, Republic of Korea, Luxembourg, Netherlands, Norway, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Turkey, and South Africa.

Lot / Code Info

UDI - (01)00643169693142(10)

Root Cause

Software design

Action Taken

A Field Corrective Action was initiated outside of the U.S. by the recalling firm's global regions on 7/4/2018.