Abbott ARCHITECT Creatine Kinase, IVD, REF 7D6341.

This recall has been terminated (originally issued July 18, 2018).

Company
Abbott Laboratories, Inc
Recall Initiated
July 18, 2018
Terminated
August 13, 2020
Recall Number
Z-1076-2019
Quantity
30 kits
Firm Location
Irving, TX
Official Source
View on FDA website ↗

Reason for Recall

The product has a stability issue which may lead to an error code indicating "Unable to calculate result, reaction check failure" and/or Quality Control results out of range low.

Distribution

Distribution was made nationwide, including Puerto Rico. There was also government and military distribution. Foreign distribution was made to Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gaza & Jericho, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Kazakhstan, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mexico, Moldova, Morocco, Netherlands, Netherlands (Curacao), New Zealand, Norway, Oman, Pakistan, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Yemen, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turkey, UAE, Uganda, United Kingdom, Uruguay, Vietnam, and Zimbabwe.

Lot / Code Info

Lot number 99633UN18, exp. 02JAN19, UDI (01)00380740101176(17)190102(10)99633UN18.

Root Cause

Material/Component Contamination

Action Taken

Recall letters were issued via FedEx overnight priority mail on 7/18/2018.