Medtronic CareLink Encore 29901 Programmer

Company
Medtronic Inc., Cardiac Rhythm And Heart Failure (Crhf)
Recall Initiated
October 2, 2018
Recall Number
Z-0407-2019
Quantity
4072 units
Firm Location
Mounds View, MN

Reason for Recall

Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.

Distribution

Worldwide

Lot / Code Info

All Serial Numbers

Root Cause

Software design

Action Taken

The firm began disseminating the Field Corrective Action notices by hand delivery on 10/02/2018 and followed with the posting of Medtronic Security Bulletin and the Urgent Medical Device Correction customer letter online on 10/11/2018. The notice explained vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download and beginning October 11th, 2018, Medtronic will be disabling the SDN for programmer updates and will rely solely on the USB update method.