RecallDepth

TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00

Company
Mindray Ds Usa, Inc. D.B.A. Mindray North America
Recall Initiated
July 14, 2017
Terminated
December 17, 2018
Recall Number
Z-0554-2018
Quantity
203
Firm Location
Mahwah, NJ

Reason for Recall

A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.

Distribution

Distributed in 26 states: AL, CA, CO, FL, GA, IA, IN, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SD, TX, UT, WA, and WY; and Puerto Rico.

Lot / Code Info

All battery chargers with part number 115-030108-00

Root Cause

Component design/selection

Action Taken

Certified letters were sent to all affected customers on 07/14/2017, to offer a replacement of the battery charger.