TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00
- Recall Initiated
- July 14, 2017
- Terminated
- December 17, 2018
- Recall Number
- Z-0554-2018
- Quantity
- 203
- Firm Location
- Mahwah, NJ
Reason for Recall
A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.
Distribution
Distributed in 26 states: AL, CA, CO, FL, GA, IA, IN, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SD, TX, UT, WA, and WY; and Puerto Rico.
Lot / Code Info
All battery chargers with part number 115-030108-00
Root Cause
Component design/selection
Action Taken
Certified letters were sent to all affected customers on 07/14/2017, to offer a replacement of the battery charger.