The GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0620FR; 02.07.0110PSF - 02.07.0620PSF

Company
Medacta Usa Inc
Recall Initiated
October 24, 2017
Posted
December 20, 2017
Terminated
June 20, 2018
Recall Number
Z-0266-2018
Quantity
12,768 units
Firm Location
Memphis, TN

Reason for Recall

Medacta has observed a worldwide occurrence rate of post-operative insert screw back-out of approximately 0.1%. For this reason, we stress the importance of using the 3.5 Nm torque limiting driver if the screw is used, as directed in our surgical techniques.

Distribution

nationwide

Lot / Code Info

all serial numbers

Root Cause

Device Design

Action Taken

All surgeon customers shall be notified via email by October 24, 2017 of the reminder to follow the surgical technique. To ensure that surgeons receive the notification, we will send the letter via email so that we can track and ensure the message was received. If the email is returned, we would then send a personalized letter directly to the surgeon. A copy of this notification is included with the submission of this report.