TERUMO(R) Overpressure Safety Valve, REF: (a) LH130 (b) LH130J

This recall has been terminated (originally issued June 27, 2017).

Company
Terumo Cardiovascular Systems Corporation
Recall Initiated
June 27, 2017
Terminated
July 16, 2019
Recall Number
Z-0256-2018
Quantity
297,222 units
Firm Location
Elkton, MD
Official Source
View on FDA website ↗

Reason for Recall

Terumo Cardiovascular Systems (Terumo CVS) has received reports of no flow through the over-pressure safety (OPS) Valve.

Distribution

Worldwide distribution: US, Canada, Colombia, UK, Belgium, and Japan

Lot / Code Info

SAFETY ALERT Lot Codes:  (a) RH14, RK11, RK20, RL02, RM06, RM20, RN10, RP08, TD16, TE13, TE27, TF14, TF21, TG04, TG11, TG17, TH06, TH16, TK10, TK24, TL08, TL21, TM28, TP14, UA11, UA18, UA25, UC08, UC22, UD14, UD21, UD28, UE05, UE18, UF09, UG06, UK01, UL12, UM10, UM31, UP05, VA02, VC06, VD06, VE10    (b) RH14, RH21, RK18, RL15, RM01, RM22, RN24, RP01, TA19, TF05, TF27, TG19, TH20, TK24, TL01, TM12, TN09, TP03, UA20, UC15, UD22, UE04, UE18, UF02, UG13, UH11, UK15, UL06, UM10, VA02, VA30, VC27, VE03, VF01   LH130 PRODUCT CORRECTION lot codes or serial number range: TA19 through VD06, distribution date range: 08Apr2015 through 20Jan2017

Root Cause

Process change control

Action Taken

The firm sent a safety alert to its consignees on 07/27/2017 by mail. The notice requested that the consignee test the product prior to using. If the product did not operate, the consignee was asked to return it. Beginning on 12/11/2017 the firm conducted a removal by mail of any remaining product. The customer instructions are as follows: CUSTOMER INSTRUCTIONS Review this Medical Device Recall Notice. Assure that all users receive notice of this issue. Refer to the Customer Response Form to identify your product that is subject to this action. Confirm receipt of this notification by completing and returning the attached Customer Response Form to the email address or fax number indicated on the form . Terumo CVS will issue a Returned Goods Authorization upon receipt of the Customer Response Form. Return all affected products to Terumo CVS. QUESTIONS? We encourage you to contact Terumo CVS with any questions or concerns: Terumo CVS Customer Service: 1.800.521.2818 Monday- Friday, 8 a.m . - 6 p.m. ET Terumo Safety Alert Fax: 1.410.392.7183 Terumo Safety Alert Email: TCVSCustomerResponseElkton@terumomedical.com