EVOLIS Microplate System (Catalog no. 89700) with TimeLiner Data Management Software, (Catalog no. 90471A.

Company
Bio-Rad Laboratories, Inc
Recall Initiated
January 10, 2019
Recall Number
Z-1906-2019
Quantity
213 devices
Firm Location
Redmond, WA

Reason for Recall

User-induced circumstances can contribute to the EVOLIS Microplate Processor System not handling repeat testing requests/results properly and may contribute to transmitting erroneous results to the customer's Laboratory Information System (LIS).

Distribution

Distribution was nationwide to medical facilities, including to Puerto Rico. There was also government/military distribution. There was no foreign distribution.

Lot / Code Info

Software version 1.4, UDI for the software is 03610520562024.

Root Cause

Software design

Action Taken

The recalling firm issued letters dated 1/9/2019 via FedEx on 1/10/2019 to all EVOLIS customers.