Disposable 25GA Vitrectomy Cutter, REF NGP0025, RX Only, Abbott Medical Optics Inc., The product provides the item necessary to perform one vitrectomy procedure. It is used to remove vitreous from the eye. It has been designed as a disposable (single use)

Company
Johnson & Johnson Surgical Vision Inc
Recall Initiated
September 24, 2018
Recall Number
Z-1893-2019
Quantity
392 units
Firm Location
Santa Ana, CA

Reason for Recall

Packaging error, 20GA Vitrectomy Cutter may be found in a 25GA package. Use of a Vitrectomy cutter that is a different size than expected could lead to the need to alter the surgical technique, including corneal incision enlargement.

Distribution

US Distribution to states of: NC, NV, TX, NY, UT, C, MN, NY, WA, NJ, OH, NV, CA and Internationally to: Australia and Japan.

Lot / Code Info

Lot Number: 60084881 Mfg. date: 2017-09-05 Exp. date: 2020-09-05 UDI: (01)05050474534117 (17)200905(10)60084881

Root Cause

Packaging process control

Action Taken

The firm, Johnson & Johnson Vision, sent an"URGENT MEDICAL DEVICE RECALL" Letter dated September 20, 2018 overnight on September 20, 2018 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: If you have received impacted product, please immediately take the following actions: 1. Review your inventory to identify units of affected lot; 2. Stop using and remove from your inventory all affected lot; and 3. regardless if you do not have inventory, please complete the Customer Reply form and return product. If you have questions, please contact the firm at 1-877-166-4543.