Intera Achieva 1.5T IT Nova Dual, Model Number 781176. Nuclear Magnetic Resonance Imaging System

This recall has been terminated (originally issued February 22, 2019).

Company
Philips Medical Systems Nederlands Veenpluis 4-6 Best Netherlands
Recall Initiated
December 21, 2018
Posted
February 22, 2019
Terminated
November 19, 2020
Recall Number
Z-0892-2019
Quantity
11226 total
Official Source
View on FDA website ↗

Reason for Recall

There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.

Distribution

U.S. Nationwide distribution.

Lot / Code Info

All systems

Root Cause

Unknown/Undetermined by firm

Action Taken

Urgent Medical Device Correction notification letters dated December 2018 were sent to customers.