MyLab Ultrasound System; Models 6100, 6150, 6200, 6250; Probes: BL433, C 18, CA541, EC1123, EC123, IH 618, IOT332, IOT342, L 415, LA332E, LA435, LA523, LA533, SI2C41, SL2325, TRT33

Company
Esaote S.P.A. Sesto Fiorentino Via Degli Olmi 11 Sesto Fiorentino Italy
Recall Initiated
January 3, 2018
Terminated
May 30, 2019
Recall Number
Z-1138-2018
Quantity
5

Reason for Recall

The ultrasound system used in certain settings may result in overheating of the ultrasound probe head.

Distribution

US distribution in the states: AZ, FL, OK, PA, and TX.

Lot / Code Info

Software versions 12.10 - 13.60

Root Cause

Software design

Action Taken

The January 3, 2018, notification letter requests the consignee complete and return a recall response form. The letter states that Esaote customer service will contact the consignee about the software update. In the meantime, it is recommended that the consignee does not use the PA023E probe in Pulsed Wave (PW) Doppler mode and that they refrain from using the system in Elaxto mode where the patient is unconscious or sedated. The consignee is asked to check the temperature of the probe before use.