RecallDepth

Magellan Diagnostics LeadCare Ultra and LeadCare Plus test Kits. Product Usage: The LeadCare Blood Lead Test Kit is for in vitro diagnostic use only. The Test Kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and Treatment Reagent of the Test Kit are specific for lead only. Contents of this LeadCare Blood Lead Test Kit must only be used with a LeadCare Analyzer and Blood Lead Testing System.

This recall has been terminated (originally issued June 5, 2017).

Company
Magellan Diagnostics, Inc.
Recall Initiated
May 23, 2017
Posted
June 5, 2017
Terminated
July 8, 2021
Recall Number
Z-2226-2017
Quantity
6018 kits (total)= 5818 (US)and 200 (OUS)
Firm Location
North Billerica, MA
Official Source
View on FDA website ↗

Reason for Recall

Underestimates the lead concentration of venous blood samples when the sample is analyzed with the LeadCare systems

Distribution

Worldwide Distribution - US (Nationwide) Foreign: AUSTRALIA AUSTRIA BOLIVIA CANADA CHILE CHINA COLOMBIA DOMINICAN REPUB ENGLAND GERMANY INDIA UK ISRAEL ITALY JAPAN KOREA MALI MYANMAR SPAIN NEPAL NETHERLANDS NEW ZEALAND P.R. CHINA PANAMA US PERU PHILLIPINES POLAND PORTUGAL REP OF KOSOVO SOUTH AFRICA SRI LANKA USA SWEDEN SWITZERLAND TRINIDAD UNITED ARAB EMI VIETNAM

Lot / Code Info

All lot codes

Root Cause

Device Design

Action Taken

Magellan issued a Safety Notification letter from May 23-25, 2017 . The firm states following the FDA Safety Notice, published May 17, 2017, this notification is to advise your facility regarding a change to existing LeadCare product usage and labeling. Prior instructions for use included capillary and venous blood samples. However, because the LeadCare Testing Systems may underestimate blood lead levels and give inaccurate results when processing venous blood samples, the FDA recommends discontinuing the use of venous blood samples with any LeadCare Blood Lead Testing System (LeadCare, LeadCare II, LeadCare Ultra, LeadCare Plus). Capillary samples may still be used on all LeadCare prducts listed above. Users requested to complete/sign Fax Form Record to verify that you have read and understood the Safety Communication to Customers on page 1. Contact Reba Daoust, Director of QA/RA 978-248-4811 or rdaoust@magellandx.com