HexaPOD evo RT System Product Usage: The intended use of the device is the control of accurate patient positioning with assistance of a 3D Tracking System in a radiotherapy environment.

Company
Elekta, Inc.
Recall Initiated
August 18, 2016
Posted
August 30, 2016
Terminated
May 23, 2017
Recall Number
Z-2691-2016
Quantity
13
Firm Location
Atlanta, GA

Reason for Recall

Potentially unrecognized incorrect position of the treatment couch in 3D workflow, i.e. the HexaPOD has not moved fully to the 3D position.

Distribution

Worldwide Distribution - US Nationwide in the states of LA, PA, WA, and the countries of: Australia, Denmark, France, India, Italy, and Japan.

Lot / Code Info

iGuide 2.2.0

Root Cause

Nonconforming Material/Component

Action Taken

Elekta sent an Important User Notice 618-03-303-025 to affected customers on 8/18/2016. The notice identified the affected product, problem, clinical impact, solution and actions to be taken. The Notice reminds users that the Interlock Check is not possible during 3D Workflow as iGUIDE is not in use. The notice reminds users that if iGUIDE is not being used - (A) Ensure the HexaPOD is at its pre-defined 3D position before treatment, and (B) Ensure the External Inhibit LED at the EnableSwitch Board is off. For questions contact your local Elekta office.