SENSAR 1-Piece IOL, Model No. AAB00 The product is indicated for implantation for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed by extracapsular cataract extraction. The lens is intended to be placed in the capsular bag.

Company
Abbott Medical Optics Inc (Amo)
Recall Initiated
August 26, 2016
Posted
September 21, 2016
Terminated
December 23, 2016
Recall Number
Z-2860-2016
Quantity
1 unit
Firm Location
Santa Ana, CA

Reason for Recall

Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.

Distribution

Worldwide Distribution - US nationwide, PR, and Internationally to Canada, Latin America, Europe, Africa, Middle East, and Asia Pacific.

Lot / Code Info

Serial No. 5050941602

Root Cause

Process control

Action Taken

Abbott Medical Optics, Inc. sent an Urgent Field Safety Notice to all affected customers to inform them that Abbott Medical Optics (AMO) is recalling 737 IOLs due to detection of an inspection equipment malfunction that resulted in a total of 737 distributed lenses not being fully checked prior to release. The letter informs customers that this malfunction may result in the release of mislabeled IOLs. The letter informs the customers of the actions to be taken. Customers are instructed to complete Customer Reply Form and fax it to AMO Quality Assurance at 714-247-4510 or email to RegCompliOne@abbott.com. Customers with any questions related to the IOL return process are instructed to contact an AMO Customer Service Representative at 877-266-4543. Customers with product complaints or adverse events are instructed to inform AMO by calling 877-266-4543.