HemosIL Plasminogen, Part No. 0020009000, IVD. Instrumentation Laboratory Company. Kit contents: 2 x 2.5 mL Streptokinase reagent and 2 x 2 mL Chromogenic substrate

This recall has been terminated (originally issued February 1, 2016).

Company
Instrumentation Laboratory Co.
Recall Initiated
December 18, 2015
Posted
February 1, 2016
Terminated
August 30, 2017
Recall Number
Z-0736-2016
Quantity
Domestic: 108 kits, Foreign: 197 kits
Firm Location
Bedford, MA
Official Source
View on FDA website ↗

Reason for Recall

This Lot is not meeting labeled on-board instrument stability claims of 5 days.

Distribution

Worldwide Distribution - US Distribution and to the countries of : Canada, Argentina, Austria, Belgium, Brazil, China, Czech Republic, France, Germany, Hong Kong, Italy, Lithuania, Panama, Poland, Russia, Saudi Arabia, United Kingdom, and Venezuela.

Lot / Code Info

Lot No. N0455811, Expiration Date 03/31/2017

Root Cause

Device Design

Action Taken

Instrumentation Laboratory Co. sent an " Urgent Medical Device Removal" letter dated December 2015, was sent on December 18, 2015. via USPS to US and Canadian customers. The notification requires US and Canadian customers to respond using a Mandatory Response Tracking Form. A Field Safety Notice was e-mailed on December 18, 2015, to the affected IL affiliates, notifying them of the product lot removal due to a stability issue and requesting them to translate the Field Safety Notice, to notify their Competent Authorities and to provide the notification to their affected customer base and secondary distributors. For further questions, please call 1(800) 955-9525.