Zimmer UNIVERSAL Power System Loaner and Modular Electric/Battery Double Trigger Handpiece. Rx only Made in Switzerland

This recall has been terminated (originally issued December 21, 2015).

Company
Zimmer Inc.
Recall Initiated
November 10, 2015
Posted
December 21, 2015
Terminated
February 19, 2016
Recall Number
Z-0452-2016
Quantity
87
Firm Location
Warsaw, IN
Official Source
View on FDA website ↗

Reason for Recall

The firm identified a malfunction which made it possible for a handpiece to start by itself when the power source is connected.

Distribution

US Distribution to the states of : AR, CA, CO, CT, FL, GA, HI, IL, KS, KY, LA, ME, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, PA, RI, SC, TX, UT, VA, WA, WI, WV and WY.

Lot / Code Info

Model # 01-8507-400-00 & Model # 89-8507-400-00

Root Cause

Device Design

Action Taken

Zimmer sent an Urgent Medical Device Correction letter dated November 9, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Review this notification. 2. You will receive copies of notification sent directly to hospital risk managers and surgeons in your territory. As necessary, review and facilitate understanding of this notification by those entities. 3. Further identify hospital risk managers and/or surgeons in your territory who should also receive notification of this update, and supply the information to the entities you have identified. 4. If after reviewing this notification you have further questions or concerns, please call 330-364-0989 between the hours of 8 a.m. and 5 p.m. EST, Monday through Friday. Zimmer instructed their direct customers that they will upgrade the device with the new electronic during their annual preventative maintenance activities. Until the new upgrade is installed, the firm instructed their customers to follow the Instructions for Use.