RecallDepth

ReUnion TSA Shoulder System, Standard SR Humeral Head Trial, 48X18, Catalog number 5901-S-4818. Orthopedic manual surgical instrument. Howmedica Osteonics Corp. A Subsidiary of Stryker Corp. Stryker France Non-sterile Made in Cork Ireland

This recall has been terminated (originally issued August 10, 2015).

Company
Stryker Howmedica Osteonics Corp.
Recall Initiated
May 12, 2015
Posted
August 10, 2015
Terminated
October 5, 2016
Recall Number
Z-2363-2015
Quantity
21 units
Firm Location
Mahwah, NJ
Official Source
View on FDA website ↗

Reason for Recall

A package labeled as Part #5901-S-4818, Lot MAC7C14, standard humeral head trial, contained part # 5901-E-4818 Lot MAC7C14, offset humeral head trial.

Distribution

US Nationwide

Lot / Code Info

Catalog #5901-S-4818, Lot MAC7C14.

Root Cause

Process control

Action Taken

Stryker Orthopaedics notified their branches/agencies via email on 5/12/2015 and sent notification letters and product accountability forms to hospital risk management/branches on 7/15/2015 and 7/16/2015

More recalls by Stryker Howmedica Osteonics Corp.

View all recalls by this company →