Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number 0998-00-0708-01. The Panorama Network includes the Panorama Telemetry System, which acquires and monitors physiological data for ambulating patients within a defined coverage area.

This recall has been terminated (originally issued May 26, 2015).

Company
Mindray Ds Usa, Inc. D.B.A. Mindray North America
Recall Initiated
March 12, 2015
Posted
May 26, 2015
Terminated
March 28, 2017
Recall Number
Z-1675-2015
Quantity
4 units
Firm Location
Mahwah, NJ
Official Source
View on FDA website ↗

Reason for Recall

Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.).

Distribution

Worldwide Distribution - US Nationwide in the state of TX.

Lot / Code Info

Part No: 0998-00-0708-01 Software version 8.9

Root Cause

Software design

Action Taken

Mindray sent a Corrective Action letter dated March 12, 2015 to their affected customer via FedEx. The letter identified the affected product, problem and actions to be taken. For questions call 201-995-8407.