Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land marks.

This recall has been terminated (originally issued December 16, 2014).

Company
Varian Medical Systems, Inc.
Recall Initiated
April 30, 2014
Posted
December 16, 2014
Terminated
March 16, 2015
Recall Number
Z-0526-2015
Quantity
32 units
Firm Location
Palo Alto, CA
Official Source
View on FDA website ↗

Reason for Recall

Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.

Distribution

Worldwide Distribution.

Lot / Code Info

Part Number P1003290001

Root Cause

Component change control

Action Taken

Varian will work with affected customers to establish schedules that minimize impact to their clinic's treatment schedule and send Field Service Engineers with the required hardware to each affected site to make the modification.