TRx Vertical Positioner, Classification Name: Wheelchair Component - Power Positioning System.

This recall has been terminated (originally issued September 17, 2014).

Company
Motion Concepts 84 Citation Drive Units 1-7 Concord Canada
Recall Initiated
August 28, 2014
Posted
September 17, 2014
Terminated
June 3, 2015
Recall Number
Z-2657-2014
Quantity
US: 59 units
Official Source
View on FDA website ↗

Reason for Recall

Motion Concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the Vertical Position System (TRx Vertical Positioner).

Distribution

US Distribution in states of: AR, CA, CT, DE, IA, ID, IL, IN, MI, MD, MO, NE, NY, OH, PA, TN, TX, and WA.

Lot / Code Info

All units manufactured between June 2011 and May 2014: Serial Numbers 28583, 28597, 28604, 28607, 28629, 30334, 028380, U001050, U001083, U001544, U002020, U002326, U002386, U002403, U002411, U002489, U002502, U002505, U002509, U002510, U002519, U002716, U002922, U003577, U004011, U004018, U004073, U004079, U004145, U004214, U004400, U004404, U004472, U004485, U006042, U006234, U006235, U006260, U006295, U006296, U006353, U006378, U007126, U007138, U007163, U007239, U007245, U007279, U007288, U007407, U008556, U008631, U008698, U008840, U008896, U009548, U009569, U009579, U009707, U009718.

Root Cause

Device Design

Action Taken

URGENT RECALL NOTIFICATION Letters (dated 8/26/2014) were sent to customers on 8/26/14 via Federal Express. Motion Concepts is requesting that all customers that have purchased a Motion Concepts TRx Vertical Positioner system in between June 2011 - May 2014 to make arrangements for replacement of the knee support mount in accordance with the instructions enclosed in the letter. For questions, please contact us at 905-695-0134.