CONMED ADULT R2 Multifunction Electrodes, REF/Catalog Number 3115-1751, Direct Connect to Philips Heartstart, Rx ONLY --- ASSEMBLED IN MEXICO -Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring applications. This product is a single-use, disposable device.

Company
Conmed Corporation
Recall Initiated
November 6, 2014
Posted
December 4, 2014
Terminated
October 30, 2017
Recall Number
Z-0462-2015
Quantity
174,610 (168,090 US and 6,520 OUS) in total.
Firm Location
Utica, NY

Reason for Recall

Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.

Distribution

Worldwide Distribution - USA (nationwide) Puerto Rico, and Internationally to Bosnia, Herzegovina, Brazil, Canada, Great Britain, Israel, Italy, Republic of Korea, Norway, Portugal, Saudi Arabia, and Spain.

Lot / Code Info

All Lot Codes

Root Cause

Device Design

Action Taken

ConMed sent an Urgent Device Correction letter dated November 16, 2014, to all affected consignees via UPS Next Day Air. The letter instructed consignees to notify all end users of the correction. For questions or additional information, consignees were instructed to contact ConMed at 1-727-399-5276 or by e-mail at multifunctionelectrodes@conmed.com. International notices were sent via UPS Priority or equivalent by November 19, 2014.