DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 1019397 (all other countries) --- Carestream Health, Inc. The DRX-Revolution Mobile X-Ray System is a mobile imaging system that incorporates a self contained x-ray generator, image receptor, imaging display and software for acquiring medical diagnostic images outside of a standard x-ray room. The DRX-Revolution Mobile X-Ray System has a motor for driving the system so that the operator does not have to push the system, but can walk behind the system while it is moving from one location to another.

Company
Carestream Health, Inc.
Recall Initiated
November 25, 2014
Posted
December 16, 2014
Terminated
June 6, 2016
Recall Number
Z-0800-2015
Quantity
629 units (Domestic: 358 units; Foreign: 271 units)
Firm Location
Rochester, NY

Reason for Recall

While servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue related to the DRX-Revolution Mobile X-Ray System in which bolts that secure the x-ray tube assembly/boom to the device mounting block may break. It has been determined that this issue is a result of an improperly assembled part which holds the x-ray

Distribution

Worldwide Distribution: US (nationwide) and countries of: AUSTRALIA, BELGIUM, CANADA, CHILE, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, ITALY, JAPAN, KOREA, KUWAIT, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, and UNITED KINGDOM.

Lot / Code Info

Serial Numbers 101 to 764

Root Cause

Device Design

Action Taken

The firm, Carestream, sent an "URGENT MEDICAL DEVICE RECALL" letter dated 11/20/2014 to US Consignees/customers via Certified Mail on 11/25/2014. Foreign accounts will be notified as soon as possible. The letter described the product, problem and actions to taken. The customers were instructed if they observe a loosened boom to follow the procedure for turning the system off; stop using the device and contact Carestream Customer Care Center in the U.S. at 1-800-328-2910, available 7 days per week on a 24 hour basis; Outside of the U.S., call your local Service support number. Carestream Health Service representative will contact all customers to schedule a visit to inspect the DRX-Revolution Mobile X-Ray System. If a repair is required it will be completed by a Carestream Field Engineer. If you have any questions, contact the Regulatory Affairs Manager at 585-627-8230 or Carestream Customer Care Center in the U.S. at 1-800-328-2910, available 7 days per week on a 24 hour basis.