IntelliFill i.v. Pharmacy Compounding Device (PCD). Intended to fill syringes from 0.5mL up to 11.5mL using mixing instructions from its formulary to fill the orders.

This recall has been terminated (originally issued December 29, 2013).

Company
Baxter Healthcare Corp.
Recall Initiated
November 8, 2013
Posted
December 29, 2013
Terminated
March 23, 2015
Recall Number
Z-0607-2014
Quantity
28 devices
Firm Location
Deerfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Following a retrospective review for Baxter Daytona Complaint CR-954, it was identified that the IntelliFill i.v. Pharmacy Compounding Device (PCD) (product code: IFIV) has the potential to continue to run and fill syringes with the HEPA filtration disabled.

Distribution

Nationwide Distribution including the states of CA, CO, CT, FL, IL, LA, MA, MD, MI, NC, NE, NY, OH, OR, PA, TN, TX, and WV.

Lot / Code Info

Lot # 123, 125R, 127, 129, 145, 146R, 147R, 148, 215R, 500, 501, 502, 503, 504, 505, 506, 508, 510, 511R, 512, 514, 515, 516R, 517, 518, 519, and 520.

Root Cause

Nonconforming Material/Component

Action Taken

A Service Bulletin was sent to all IntelliFill i. v. customers February 22, 20 I 0 by USPS mail. The Service Bulletin informs customers that IntelliFill i.v. operators are to be trained to listen for the running HEPA filtration fan motor and observe/feel for airflow coming from the automation deck. The field service engineer dispatched to the site replaced the damaged sensor on 11-Feb-2010 and a Service Bulletin was sent to all customers regarding the function and use of the HEPA filter on 22-Feb-2010. A new HEPA sensor was designed and installed into all current IntelliFill i.v. devices between 21-July-2011 to 9-Feb-2012.