Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004

This recall has been terminated (originally issued December 24, 2013).

Company
Data Innovations, Inc.
Recall Initiated
November 6, 2013
Posted
December 24, 2013
Terminated
December 5, 2014
Recall Number
Z-0571-2014
Quantity
24 systems
Firm Location
South Burlington, VT
Official Source
View on FDA website ↗

Reason for Recall

When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error

Distribution

Nationwide Canada and Britain

Lot / Code Info

Software alll versions prior to: v7.00.0006 and v8.00. Includes: v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004

Root Cause

Software design

Action Taken

Data Innovation notified consignees by Important Medical Device Recall Notification dated 11/5/13 to all list serv recipients of the problem and posted critical alert on Customer Web Portal. A certified letter was issued to impacted customers and consignees if no response was recieved from e-mail.