Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
This recall has been terminated (originally issued December 24, 2013).
- Company
- Data Innovations, Inc.
- Recall Initiated
- November 6, 2013
- Posted
- December 24, 2013
- Terminated
- December 5, 2014
- Recall Number
- Z-0571-2014
- Quantity
- 24 systems
- Firm Location
- South Burlington, VT
- Official Source
- View on FDA website ↗
Reason for Recall
When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error
Distribution
Nationwide Canada and Britain
Lot / Code Info
Software alll versions prior to: v7.00.0006 and v8.00. Includes: v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
Root Cause
Software design
Action Taken
Data Innovation notified consignees by Important Medical Device Recall Notification dated 11/5/13 to all list serv recipients of the problem and posted critical alert on Customer Web Portal. A certified letter was issued to impacted customers and consignees if no response was recieved from e-mail.