RecallDepth

AlboGraft Woven with Collagen, Straight 30x16 Catalog Number: ATC3016 AlboGraft Vascular Grafts are made of synthetic material and designed to replace sections of damaged or malfunctioning arteries.

Company
Lemaitre Vascular, Inc.
Recall Initiated
June 17, 2013
Posted
July 22, 2013
Terminated
August 15, 2016
Recall Number
Z-1767-2013
Quantity
1 unit
Firm Location
Burlington, MA

Reason for Recall

Blood blushing/leaking from the surface of the graft after implantation

Distribution

Worldwide distribution: USA including PA and countries of: BRAZIL, BULGARIA, FRANCE, GEORGIA, GERMANY, GREECE, ITALY, LEBANON, NETHERLAND, POLAND, SPAIN, and TURKEY.

Lot / Code Info

Lot number/Expiration Date  56890A 2015-04

Root Cause

Process control

Action Taken

LeMaitre Vascular, Inc.sent an Urgent Field Safety Notice daed June 19, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to identify and return the affected device to MeMaitre Vascular which they will replace with a new unit. Customers with questions were instructed to call 781-221-2266, ext. 183. For questions regarding this recall call 781-425-1670, ext 108.

More recalls by Lemaitre Vascular, Inc.