STERRAD 200 System Carriage, P/N 10208. Used as a tool to expedite loading and unloading processed items from a STERRAD 200 sterilizer.

This recall has been terminated (originally issued March 26, 2013).

Company
Advanced Sterilization Products
Recall Initiated
March 13, 2012
Posted
March 26, 2013
Terminated
May 28, 2014
Recall Number
Z-0998-2013
Quantity
2,457 units
Firm Location
Irvine, CA
Official Source
View on FDA website ↗

Reason for Recall

Advanced Sterilization Products (ASP) is recalling the STERRAD 200 System Carriage because it can potentially become dislodged when an excessive load is applied.

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Egypt, Germany, Italy, France, Saudi Arabia, UAE, Japan, Singapore, China, Thailand, Hong Kong, Hungary, Israel, Slovania, Poland, Czech Republic, Taiwan, Sweden, Greece, Korea, Ireland, and United Kingdom.

Root Cause

Device Design

Action Taken

Advanced Sterilization Products (ASP) sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated March 13, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The letter informed the customers that ASP recently determined that the push bar located on the STERRAD 200 System Carriage can potentially become dislodged when an excessive load is applied. Customers with questions are instructed to contact Stericycle at (877) 931-0642.