MediChoice Rayon Tipped OB/GYN Applicator, 8 inch, Plastic Shaft, Non-Sterile, Item Numbers WOD5001 and WOD5002. Product Usage: Usage: An absorbent tipped applicator is a device intended for medical purposes that consists of an absorbent swab on a wooden, paper, or plastic stick. The device is used to apply medications to, or to take specimens from, a patient.

This recall has been terminated (originally issued May 1, 2013).

Company
Owens & Minor Distribution, Inc.
Recall Initiated
March 26, 2013
Posted
May 1, 2013
Terminated
October 31, 2014
Recall Number
Z-1192-2013
Quantity
4298.5 cases
Firm Location
Mechanicsville, VA
Official Source
View on FDA website ↗

Reason for Recall

Owens & Minor Distribution, Inc. is conducting a market withdrawal of MediChoice Applicator OB-GYN 8 inch product that does not meet our quality standards because the rayon may loosen or unravel from the applicator during use.

Distribution

US Nationwide Distribution.

Lot / Code Info

Item # Lot#  WOD5001 1206DG13A  WOD5001 1206DG13B  WOD5001 1207DG13A  WOD5001 1208DG13A  WOD5001 1209DG13A  WOD5001 1210DG13A/B  WOD5001 1211DG13A  WOD5001 1212DG13A  WOD5001 1301DG13A  WOD5001 1303DG13A    WOD5002 1206DG14A  WOD5002 1208DG14A  WOD5002 1209DG14A  WOD5002 1211DG14A   Expiration Dates: 2017-05, 2017-06, 2017-06, 2017-07, 2017-08, 2017-09, 2017-10, 2017-11, 2018-02.

Root Cause

Process change control

Action Taken

Owens&Minor sent an Urgent: Medical Device Recall letter to the first set of customers on March 26, 2013 and to their second set of customers on April 9, 2013. The letter identified the affect product, problem and actions to be taken. Customers were instructed to examine their inventory and quarantine all affected products, contact service representative to discuss whether affected products should be returned or destroyed, and complete and return the Recall Reply Form. For questions call 804-723-7417.