Flexible Grip which is part of the Synthes Universal Nail System. Indicated for the conventional (non-locking) technique of treating stable diaphyseal fractures.

Company
Synthes Usa (Hq), Inc.
Recall Initiated
January 11, 2013
Posted
April 5, 2013
Terminated
August 31, 2015
Recall Number
Z-1062-2013
Quantity
328
Firm Location
West Chester, PA

Reason for Recall

Synthes is initiating a voluntary recall of the Flexible Grip (Part number 355.28) which is part of the Universal Nail System, due to the potential lack of the ability to thoroughly clean the instrument.

Distribution

Nationwide Distribution including the states of AZ, CA, CO, FL, IA, ID, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NV, NY, OH, PA, RI, TN, TX, UT, VT, WA, WI, and WV.

Lot / Code Info

Part number 355.28, all lot numbers

Root Cause

Device Design

Action Taken

Synthes sent an "URGENT NOTICE: MEDICAL DEVICE RECALL" dated January 11, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The letter requested consignees examine inventory, remove the recalled products from use and return them to Synthes. If you have any questions, contact the firm at 610-719-5450.