Depuy Spine Side Fire Needles 11G X 4IN, part of the Confidence Spinal Cement System Product Code: 283904411. The Confidence Side Fire Needle is used to place bone cement in an intra-vertebral fashion at a desired location. the cement extrudes fro the needle through a port on the side of the needle.

This recall has been terminated (originally issued July 3, 2012).

Company
Depuy Mitek, Inc., A Johnson & Johnson Co.
Recall Initiated
May 4, 2012
Posted
July 3, 2012
Terminated
August 4, 2016
Recall Number
Z-1938-2012
Quantity
704 units
Firm Location
Raynham, MA
Official Source
View on FDA website ↗

Reason for Recall

Additional Instructions for Use and Surgical Technique for the Confidence Needle due to reports of breakage with the side fire needles

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Canada, Austria, China, Croatia, Czech Republic, France, Germany, Greece, Hungary, Italy, Lebanon, Netherlands, New Zealand, Poland, Russian Federation, Singapore, South Africa, Spain, Switzerland, Turkey, UAE, and UK.

Lot / Code Info

All lots

Root Cause

Labeling design

Action Taken

Depuy Spine sent a "CORRECTION NOTICE" to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Field Correction Notification Form was attached for customers to complete and return via fax to 508-828-3070. For questions about Confidence Needles, please contact your local DePuy Spine Sales Representative. If you need assistance in locating your local Sales Representative please contact JJHCS CIS @ 1 -800-255 -2500 .