RecallDepth

Symbia E Series SPECT System; Single-Photon Emission Computed Tomography. The Siemens Symbia E system is a high performance dual detector variable angle system with flexible patient positioning for planar imaging, whole-body scans, and SPECT studies.

Company
Siemens Medical Solutions Usa, Inc.
Recall Initiated
May 15, 2012
Posted
May 30, 2012
Terminated
December 23, 2016
Recall Number
Z-1686-2012
Quantity
9 units
Firm Location
Hoffman Estates, IL

Reason for Recall

Siemens has identified a potential patient safety issue for e.cam and Symbia E camera systems that have had the radial drive motor assembly replaced during a Service action. There is a remote chance that during the motor replacement the radial drive gear may have been replaced incorrectly. If the gear is installed incorrectly the mechanical brake may not engage creating a patient hazard if the sys

Distribution

Worldwide Distribution, US (Nationwide) and the countries of Australia, Belgium, Canada, China, France, Germany, Italy, Japan, Republic of Korea, Mexico, Netherlands, Portugal, Spain, Taiwan and the United Kingdom.

Lot / Code Info

Material number 10275879, serial numbers 1028, 1034, 1062, 1098, 1106, 1134, 1138, 1205 and 1283.

Root Cause

Employee error

Action Taken

On May 15, 2012, Siemens Medical Solutions USA, Inc. sent an "Urgent Field Correction Recall" letter dated April 11, 2012 to all affected customers. The letters informed customers of the potential patient safety issue for e.cam and Symbia E camera systems that have had the radial drive motor assembly replaced during a Service action. Customers were informed that a Siemens Service Engineer will be inspecting the radial drive gear on their camera system to ensure it is functioning correctly. For questions or assistance they were instructed to contact Siemens Medical Solutions USA, Inc. at 800-888-7436.

More recalls by Siemens Medical Solutions Usa, Inc.

View all recalls by this company →