Synthes ProDisc-C 2.0mm Milling Bit, Synthes 2.0mm Milling Bit HXC, Synthes 2.0mm Milling Bit with 2.35mm x 33.5mm Shaft, and Synthes 2.0mm Milling Bit with Flat Coupling. Spinal surgical instrument: The Synthes ProDisc-C 2.0mm Milling Bits are used as a part of the milling option for keel preparation of the vertebral bodies. The milling bits are coupled to a power tool and interface with the milling guide, which is connected to a trial implant for the keel preparation step.

This recall has been terminated (originally issued May 29, 2014).

Company
Synthes, Inc.
Recall Initiated
April 29, 2014
Posted
May 29, 2014
Terminated
May 19, 2016
Recall Number
Z-1669-2014
Quantity
6690
Firm Location
West Chester, PA
Official Source
View on FDA website ↗

Reason for Recall

Certain lots of the ProDisc-C 2.0mm Milling Bits have an increased potential to break during use.

Distribution

US Nationwide Distribution

Lot / Code Info

part number: 03.820.161s, with lot numbers: 5742507,5755970,5755971,5755972, 5755973,5755974,5828298,6336115, 6338099,6338100,6340029,6340030, 6344199,6344200,6363601,6385892, 6407046,6431260,6431261,6460402, 6503360,6526127,6571070,6571073, 6571074,6601985,6636946,6636947, 6636948,6653819,6653820,6670383, 6691341,6691342,6718933,6718934, 6718935,6740611,6794651,6794652, 6794653,6794654,6794655  part number: 03.820.1638 with lot numbers: 5742506,5755978,5755979,5755980, 5755981,5755982,5828299,5828300, 6320586,6336108,6336110,6336111, 6340033,6344201,6344202,6363602, 6385893,6431262,6431263,6460403, 6503361,6571075,6601986,6601987, 6636949,6636950,6636951,6653810, 6653818,6670382,6691343,6691344, 6718938,6740615,6740616,6745821, 6770474,6770475,6770476,6770477, 6770478,6794657,6794658  part number: 03.820.167S with lot numbers:  5755983,5755984,5755987,5755988, 5755989,5755985,5755986,5755990, 5828302,6338101,6338102,6338103, 6338104,6338106,6338107,6338108, 6338109,6344205,6471089,6670381, 6718939,6718940,6718941,6843907, 6843906,6856252,6856253,6867662, 6867661,6901712,6901713,6919726, 6919727,6919728,6901710,6901711, 6921717,6921718,6940357,6960192, 6960193,6983033,7006496,7055427  part number: 03.820.169S with lot numbers: 6066520,6113671,6113633,6159883, 6180175,6338068,6338098,6344206, 6363603,6407047,6431264,6460404, 6526128,6571078,6670384,6745820, 6770479,6770480

Root Cause

Process control

Action Taken

DePuy Synthes sent a Recall letter, dated April 29, 2014, to affected customers. The letter identified the affected customers, problem and actions to be taken. The letter instructed customers to discontinue use and distribution of the device and to return affected devices to Synthes. For questions call 610-719-5450.