Model Kodak 2200 Intraoral X-Ray System, Catalog Numbers 5303177, 5303169, 5303151, XR-5154323 -- Carestream Health, Inc. Made in France -- --- COMMON/USUAL NAME: Dental X-Ray Systems -- CLASSIFICATION NAME: Unit, X-Ray, Extraoral with timer --- There are four major components: x-ray tube, x-ray generator, collimator and image receptor. These systems are permanently installed diagnostic dental x-ray systems for generation of x-rays for examination of the teeth, jaw and oral structures.

Company
Carestream Health, Inc.
Recall Initiated
January 24, 2014
Posted
March 4, 2014
Terminated
June 18, 2015
Recall Number
Z-1128-2014
Quantity
Domestic: 347 units (total for both)
Firm Location
Rochester, NY

Reason for Recall

Carestream received problem reports relating to apparent early failure of Kodak 2X00 Intraoral X-ray device arms leading to arm breakage.

Distribution

Worldwide distribution. USA (nationwide) and the country of Canada.

Lot / Code Info

Serial Numbers starting with WE, WF, WG, WH, WI, WJ, WK, WL, XA, XB, XC, XD, XE, XF, XG, XH, XI, XJ, XK, XL, YA, YB, YC, YD

Root Cause

Component design/selection

Action Taken

Urgent Medical Device Recall Letters (dated 1/24/2014) were sent to the consignees on 1/24/2014 via Certified Mail, Return Receipt Requested.