Artis Zee Angiographic x-ray systems.
This recall has been terminated (originally issued March 31, 2014).
- Recall Initiated
- February 7, 2014
- Posted
- March 31, 2014
- Terminated
- March 16, 2015
- Recall Number
- Z-1309-2014
- Quantity
- 9
- Firm Location
- Malvern, PA
- Official Source
- View on FDA website ↗
Reason for Recall
An inappropriate cable connection was assembled for certain Artis zee systems. which may cause a system failure and possibly lead to a restricted maneuverability of the gantry system.
Distribution
MN, WI, PA, MI, AL, OH, NC, AR, and NJ.
Lot / Code Info
Model numbers: 10094135, 10094137, 10094139, 10094141, 10280959.
Root Cause
Device Design
Action Taken
A recall notification letter dated February 7, 2014, was sent to End Users.