HAMILTON-C2 Ventilator: Manufactured by Hamilton Medical AG CH-7402 Bonaduz, Switzerland; Imported and distributed by Hamilton Medical, Inc. Reno, NV 89502 The HAMILTON C2 ventilator is intended to provide positive pressure ventilatory support to adults, pediatrics, infants and neonates depending on its configuration. Intended areas of use: In the intensive care ward or in the recovery room; During transfer of ventilated patients within the hospital.

Company
Hamilton Medical Inc
Recall Initiated
September 26, 2013
Posted
November 6, 2013
Terminated
March 29, 2018
Recall Number
Z-0170-2014
Quantity
316 devices distributed in the US.
Firm Location
Reno, NV

Reason for Recall

Manufacturer defined limits for improperly functioning loudspeakers of the HAMILTON C2 alarm system were exceeded.

Distribution

Nationwide distribution: USA including Puerto Rico and the states of IL, MT, IN, ND, AZ, MT, KY, NC, OH, MA, MO, VA, CA, NV, TX, WA, NY, GA, SC,MN, WY, PA, ND, UT, NE, CO and MI.

Lot / Code Info

Hamilton-C2; Serial numbers from 1000-7100.

Root Cause

Device Design

Action Taken

Hamilton Medical sent an Medical Device Safety Alert and Corrective Action letter dated September 27, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The corrective action does not require any action by the operator. In the case of an alarm situation as described in the letter, users should follow the instruction for use (Section 8 - responding to alarms in the operator manual) and have the ventilator serviced by an authorized service technician. An addendum to the Operators Manual as a result of the field correction was provided to customers with the Safety Alert Letter. Manufacture will develop and provision a test protocol for the service tech which test the proper functioning of the loudspeaker during the annual service inspection. We appreciate your support in this matter and sincerely regret any inconveniences that this action may cause you. We consider this action as necessary to ensure that our customers receive only safe and effective products with high quality. Further questions please call (775) 426-6331 Ext. 210