Cordis, a Johnson and Johnson Company, ExoSeal(TM) 6F Vascular Closure Device, cat. No EX500, EX600, and EX700. Used for femoral artery puncture site closure, reducing times to hemostasis and ambulation in patients who have undergone diagnostic or interventional procedures.

This recall has been terminated (originally issued January 25, 2013).

Company
Cordis Corporation
Recall Initiated
October 29, 2012
Posted
January 25, 2013
Terminated
November 6, 2013
Recall Number
Z-0740-2013
Quantity
98,675 units distributed in US.
Firm Location
Miami Lakes, FL
Official Source
View on FDA website ↗

Reason for Recall

Affected lots of Cordis ExoSeal (TM) VCD have been sterilized using a process that did not achieve the sterility assurance level that is required by the standards of this class of product. Model Numbers EX500, EX600, and EX700.

Distribution

US (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and Hawaii.

Lot / Code Info

Model EX500: Lot #'s: 15410906, 15411659, 15424968, 15429098, 15429879, 15437509, 15437958, 15444208, 15449422, 15453540, 15456567, 15456568, 15462779, 15462780, 15469984, 15479199, 15481341, 15485491, 15489970, 15501955, 15505186, 15578709, 15589368, 15586530, 15602199, 15603369, 15608590, 15610801, 15617548, 15630141, 15630171, 15639388, 15642692, 15648662, 15650923, 15656441, 15664083, 15670351, 15670768.  Model EX600: Lot #'s: 15408606, 15409332, 15410622, 15410623, 15426890, 15430798, 15433095, 15433098, 15441594, 15443327, 15446063, 15448005, 15451542, 15452987, 15455770, 15455771, 15459366, 15459367, 15461179, 15461781, 15463296, 15463297, 15464298, 15464299, 15466362, 15472044, 15475128, 15475795, 15479644, 15484135, 15487458, 15491737, 15496311, 15497038, 15500078, 15500709, 15503015, 15504223, 15504224, 15504748, 15506926, 15508012, 15508353, 15508856, 15510541, 15512713, 15518232, 15540529, 15545067, 15548555, 15549324, 15550196, 15550197, 15557284, 15561545, 15561546, 15566744, 15572185, 15573480, 15579468, 15585121, 15587184, 15596793, 15601234, 15601753, 15604477, 15605089, 15609335, 15613282, 15617549, 15625033, 15627316, 15635233, 15637262, 15643624, 15644116, 15644789, 15644790, 15649759, 15663330.  Model EX700 Lot #'s: 15466361, 15471271, 15471272, 15476317, 15476743, 15482155, 15483371, 15486224, 15490397, 15535011, 15540530.

Root Cause

Material/Component Contamination

Action Taken

Cordis Corporation sent an "Urgent Medical Device Recall (Removal ) letter dated October 29, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to immediately identify and set aside all product listed in a manner that ensures the affected product will not be used. A product recall response form was attached for customers to review, complete, sign and return the enclosed acknowledgement form directly to Cordis. For further questions please call (786) 313-2000.