Acrobat 3000 (AC3000) Spring Arm as support arm for iLED operating lights, TRUMPF Medizinsysteme GmbH + Co KG 82178 Puchheim. The TRUMPF surgical lights are intended to locally illuminate an operating or examination area of the patients body with a high intensity light.

Company
Trumpf Medical Systems, Inc.
Recall Initiated
May 29, 2012
Posted
September 5, 2012
Terminated
April 11, 2013
Recall Number
Z-2311-2012
Quantity
402 units
Firm Location
Charleston, SC

Reason for Recall

Cracks can arise on the welded seam on the rear joint.

Distribution

Worldwide Distribution - USA including CA, OH, OK, NY, PA, TX, WY and the country of Canada

Lot / Code Info

I.D. on nameplate: FA AC3 or FA AC3000; Material #1920632, Serial #011201602557 to 011201602588; Serial #011203602617 to 011202602621; Serial #011203602676 to 011203602695; Serial #011203602756 to 011203602775; Serial #011203602776 to 011203602795; Serial #011204602798 to 011204602817; Serial #011204602861 to 011204602880; Serial #011204602970 to 011204602989; Serial #011205603032 to 011205603081; Serial #011206603118 to 011206603167; Serial #011206603168 to 011206603217; Serial #011207603303 to 011207603314; Serial #011208603316 to 011205603333; Serial #011208603361 - Serial #011209603431 to 011209603462; Serial #011210603572 to 011210603597 and Material #1851710, Serial #3705776 to 3705777 and Serial #4105778 to 4105781.

Root Cause

Process control

Action Taken

TRUMPF sent an Urgent Safety Information letter dated February 4, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to ensure that within their organization, all users of the affected product be informed about the Urgent Safety Information. If affected product was further distributed, customers should forward a copy of the information or inform the necessary parties. Customers were asked to confirm receipt of the Urgent Safety Information by completing and returning the confirmation of receipt. For questions regarding this recall call 888-474-9359.