Philips SureSigns VSi/VS2+ patient monitors Model Part Numbers VSi/ 863275, 863276, 863277, VS2+ 863278, 863279. The SureSigns VS2+ vital signs monitor is for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility. The SureSigns VSi vital signs monitor is for measurement of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility

This recall has been terminated (originally issued August 30, 2012).

Company
Philips Healthcare Inc.
Recall Initiated
August 15, 2012
Posted
August 30, 2012
Terminated
March 13, 2014
Recall Number
Z-2303-2012
Quantity
2314 units
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

Philips SureSigns VSi/VS2+ patient monitors may not be properly secured and may fall from the wall mount

Distribution

Worldwide Distribution-USA (nationwide) and the countries of AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, DENMARK, EL SALVADOR, FRANCE, GERMANY, ICELAND, INDIA, IRELAND ISRAEL, ITALY, JORDAN, KENYA, LEBANON, MALAYSIA,NAMIBIA NETHERLANDS, NORWAY, POLAND, PORTUGAL, RWANDA SAUDI ARABIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN SWITZERLAND, UNITED ARAB EMIRATES, and UNITED KINGDOM.

Lot / Code Info

Serial Number Range CN14800101-CN14801360  CN21201362-CN21202440, CN21202525, SP13701043, USPP101007

Root Cause

Process design

Action Taken

Philips sent an "URGENT- MEDICAL DEVICE CORRECTION" letter dated August 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Customers using the VSi/VS2+ with the GCX wall mount are instructed to complete and fax a Customer Reply Wall Mount Order Card to the Business Unit or Key Market indicating the number of wall mounts in use. Upon receipt of the Customer Reply Wall Mount Order Card, Philips will send a replacement of the mounting hardware. Customers should contact their local Philips representative for questions regarding this recall.