PrepStain Syringing Pipettes 240 (Catalog #490517)

This recall has been terminated (originally issued June 15, 2012).

Company
Tripath Imaging, Inc.
Recall Initiated
March 22, 2012
Posted
June 15, 2012
Terminated
December 3, 2012
Recall Number
Z-1811-2012
Quantity
46,112,784 for all products in Recall Event
Firm Location
Burlington, NC
Official Source
View on FDA website ↗

Reason for Recall

The PrepStain Syringing Pipettes may not dispense the minimum volume of 7.5 mL of the sample due to leaking. This event resulted in an increase in complaints of leaking syringing pipettes

Distribution

Worldwide Distribution.

Lot / Code Info

100901a, 100902f, 100902f, 100903d, 100908c, 100910c, 100914f, 100916d, 100921d, 100928c, 101005a, 101007d, 101008c, 101012c, 101013a, 101015c, 101020d, 101022c, 101027a, 101028c, 101029d, 101103b, 101105h, 101110f, 101111c, 101117p, 101118n,  101119c, 101130p, 101201o, 101207o, 101208p, 101210c, 101213n, 101214b, 101214c,  101217b, 101220q, 101221g, 110104b, 110106a, 110107c, 110110d, 110112n, 110118b,  110128i, 110203i, 110210r

Root Cause

Component design/selection

Action Taken

BD Diagnostics sent a "BD PrepMate Syringing Pipette" letter dated March 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The firm encouraged customers to refer to the Operator's Manual and inspect the centrifuge tubes to verify that the correct amount of sample has been transferred. Customers experiencing this issue were instructed to discard the pipettes and contact BD Diagnostic Technical Support for a replacement. Contact the firm at 1-877-822-7771 for questions regarding this notice.