Repose G2 Bone Screw, Sterile, REF 76310200, Lot # 62450200. Distributed by Medtronic Xomed, Inc., Jacksonville, Florida 32216 U.S.A.

Company
Medtronic Xomed, Inc.
Recall Initiated
July 27, 2009
Posted
December 15, 2009
Terminated
March 15, 2010
Recall Number
Z-0536-2010
Quantity
13
Firm Location
Jacksonville, FL

Reason for Recall

Medtronic Xomed, Inc. is recalling the Repose Bone Screw System product and Bone Screw product due to improper labeling. The labeling for the Repose System was inadvertently placed on the Repose Screw and vice versa.

Distribution

9 consignees in the continental United States.

Lot / Code Info

Lot # 62450200

Root Cause

Labeling mix-ups

Action Taken

Firm notified customers by phone on 7/27/2009. Firm asked customers to check inventories for the affected products listed on the enclosed Checklist. 1. Fill-in the "quantity on-hand "column on the checklist. 2. If you have any this product lot, contact Medtronic ENT Customer Service at 800-874-5797 (select Option 1 for Customer Service) and ask for the REPOSE Field Action Contact Person for help to arrange for returns and replacements or credit. 3. Fax your Checklist to Medtronic ENT at the number provided. 4. Return products clearly marked.