RecallDepth

Pointe Scientific Liquid ALT (SGPT) reagent set for the quantitative determination of alanine aminotransferase in serum, 625 ml, Catalog no. A7526-625, in vitro diagnostic, Pointe Scientific, Inc.; Canton, MI 48188

This recall has been terminated (originally issued May 22, 2008).

Company
Pointe Scientific, Inc.
Recall Initiated
December 14, 2007
Posted
May 22, 2008
Terminated
September 15, 2008
Recall Number
Z-1112-2008
Quantity
955 for all products.
Firm Location
Canton, MI
Official Source
View on FDA website ↗

Reason for Recall

Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms.

Distribution

Worldwide Distribution - USA, Bolivia, Canada, Chile, China, Cyprus, Greece, Indonesia, Israel, Korea, Mexico, Pakistan, Panama, Peru, Philippines, Republic of Yemen, Russia, South Africa, Trinidad, Ukraine and West Indies.

Lot / Code Info

Lots 704601-068 and 704601-123.

Root Cause

Process control

Action Taken

Pointe Scientific notified consignees via Urgent Medical Device Recall letters dated 12/14/07, instructed to dispose of the product and to send back a reply form to request replacement product.

More recalls by Pointe Scientific, Inc.

View all recalls by this company →