american diagnostica inc. IgG Conjugate, 15mL; component of the american diagnostica, inc. IMUCLONE aPL IgG ELISA Kit, Product #649G; manufactured by Louisville APL Diagnostics, Inc., Seabrook, TX 77586.
This recall has been terminated (originally issued July 24, 2007).
- Company
- Louisville Apl Diagnostics Inc
- Recall Initiated
- August 10, 2004
- Posted
- July 24, 2007
- Terminated
- November 27, 2007
- Recall Number
- Z-1104-2007
- Quantity
- 45
- Firm Location
- Seabrook, TX
- Official Source
- View on FDA website ↗
Reason for Recall
Product mislabeled; conjugate IgG placed in bottles labeled as conjugate IgM and vice versa.
Distribution
Product distributed to a single direct account in CT.
Lot / Code Info
Lot #040414.
Root Cause
Other
Action Taken
Firm notified consignee of recall via phone on 08/10/04. Consignee asked to return conjugates for properly labeled replacements.