Invasive Blood Pressure Interface Cables: a) Medrad 9500 Multigas monitor for MRI, Invasive Blood Pressure Interface Cable (Part # 3006314) for use with Abbott/Hospira IBP Transducer b) Invasive Blood Pressure Interface Cable (Part # 3008772) for use with Braun IBP Transducers. The product is an accessory to patient physiological monitor.

This recall has been terminated (originally issued December 15, 2006).

Company
Medrad Inc
Recall Initiated
September 29, 2006
Posted
December 15, 2006
Terminated
June 20, 2007
Recall Number
Z-0258-2007
Quantity
80 units
Firm Location
Indianola, PA
Official Source
View on FDA website ↗

Reason for Recall

Interference problems: certain invasive blood pressure transducers connected to these cables are susceptible to interference if exposed to static magnetic field strengths exceeding 300 gauss. In certain orientations, the transducer causes the invasive blood pressure readings on the physiological monitor to display inaccurate readings.

Distribution

Worldwide distribution ---- including USA state of CA, CO, HI, IL, IN, MA, MD, NJ, NY, OH, PA, TX, and WA and countries of The Netherlands, Denmark, Australia, Brazil, Italy, and Japan.

Lot / Code Info

Catalog number 3006314 and 3008772

Root Cause

Other

Action Taken

The recalling firm contacted all affected customers via telephone or letter dated November 20, 2006 to inform them of the problem and instruct them to return the cables or apply corrective labeling.