Program Memory Card, product code MCV1A.

This recall has been terminated (originally issued November 23, 2009).

Company
Surgical Devices/Covidien
Recall Initiated
September 23, 2009
Posted
November 23, 2009
Terminated
March 8, 2010
Recall Number
Z-0404-2010
Quantity
3992 units
Firm Location
North Haven, CT
Official Source
View on FDA website ↗

Reason for Recall

Lack of assurance that products were manufactured under GMPs. This action is being undertaken due to certain products potentially exhibiting compromised staple formation or cutting which in some instances may result in extended surgical time for the patient.

Distribution

The products were shipped medical facilities nationwide. The product was also shipped to distributors in Canada, Japan, Germany, Ireland, Italy, Spain, Taiwan, Belgium, Sweden, Netherlands, Israel, Portugal, India, Austria, Turkey, and Greece.

Lot / Code Info

all codes

Root Cause

Nonconforming Material/Component

Action Taken

The recalling firm issued an Urgent Medical Device Recall letter dated 9/22/09 to all customers. The letter informed customers that Covidien acquired PMI, about the problem with the products, and about the need to return any products remaining inventory. Customers were asked to return the entire inventory of Power Medical Interventions product, regardless of its expiration date. This process can be initiated by contacting Power Medical Innovations customer service at 1-866-769-3763.

More recalls by Surgical Devices/Covidien

View all recalls by this company →