AccuSure, U-100 31 Gauge, 1cc Insulin Syringe, 5/16" (8mm) Needles, Short Needle, packaged 10 syringes per package, 10 packages per carton, NDC 0603-7002-2

This recall has been terminated (originally issued June 24, 2009).

Company
Vintage Pharmaceuticals Llc, Dba Qualitest Pharmaceuticals
Recall Initiated
April 9, 2009
Posted
June 24, 2009
Terminated
November 1, 2010
Recall Number
Z-1504-2009
Quantity
559,500 syringes
Firm Location
Huntsville, AL
Official Source
View on FDA website ↗

Reason for Recall

Needles separate from the barrel of the syringes.

Distribution

Nationwide.

Lot / Code Info

Lot A07001, exp 07/2012

Root Cause

Nonconforming Material/Component

Action Taken

On 04/09/2009, Qualitest issued a press release and sent out recall notifications to their consignees by US Mail. The recall was extended to the User level. Qualitest will send a second letter to all non-responders after thirty days. Thirty days after the second letter, Qualitest will perform a recall effectiveness check by telephone on a percentage of non-responders (in accordance with FDA request).