Hill-Rom 100 Low Bed; Model No. P3930.

This recall has been terminated (originally issued September 5, 2008).

Company
Hill-Rom, Inc.
Recall Initiated
April 8, 2008
Posted
September 5, 2008
Terminated
October 13, 2010
Recall Number
Z-1603-2008
Quantity
753 beds
Firm Location
Batesville, IN
Official Source
View on FDA website ↗

Reason for Recall

The upper deck may collapse to its lowest position and the caster brakes may fracture if the bed is moved while the brakes are locked.

Distribution

Worldwide: USA and Canada.

Lot / Code Info

All units.

Root Cause

Device Design

Action Taken

Consignees were notified via letter dated April 8, 2008 to lower these beds to their lowest position, to remove the beds from use if they move more than 1/2 inch with the brakes locked and that the firm will repair these beds when fixes become available. If you have questions regarding this recall, contact Hill-Rom Technical Support at 800-445-3730.