The Pinnacle3 Radiation Therapy Planning (RTP) system and its associated software; composed of several modules including the core Pinnacle3 functionality, Syntegra (TM), P3IMRT, P3MD and AcQSim3 (TM). The Pinnacle3 RTP system is composed of a Sun UNIX workstation (or UNIX-complaint computer) running the Solaris operating system and software, and a UNIX terminal emulation package on a personal computer, which allows the user to enter patient data into the system, use that data to construct a plan for radiation therapy, and evaluate the plan- Philips Medical Systems-Model #9200-0652A-ENG, 9200-0700A-ENG, 9200-0700B-ENG, 9200-0700C-ENG (with software versions 7.6C, 8.0d and 8.0h)

This recall has been terminated (originally issued September 25, 2007).

Company
Philips Nuclear Medicine, Inc.
Recall Initiated
August 21, 2007
Posted
September 25, 2007
Terminated
December 17, 2011
Recall Number
Z-1212-2007
Quantity
1,407
Firm Location
Fitchburg, WI
Official Source
View on FDA website ↗

Reason for Recall

Incorrect Absolute Marking Coordinates when used with PQ-Series scanners due to a software malfunction

Distribution

Worldwide Distribution; USA including states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, and WI and Country of CANADA

Lot / Code Info

Pinnacle3 Version 7.6c Software Kit Model #9200-0652A-ENG;Pinnacle3 Version 8.0d Software Kit Model #9200-0700A-ENG;Pinnacle3 Version 8.0d p 1 Software Kit Model #9200-0700B-ENG;Pinnacle3 Version 8.0h Software Kit Model #9200-0700C-ENG

Root Cause

Other

Action Taken

All customers will receive an August 21, 2007, Customer Advisory Notification letter notifying them of the issue and the steps required to avoid the issue. Return receipts will be tracked. This is a short-term fix, a long term software fix will be issued shortly by the firm.