OEC 9900 Elite image-intensified fluoroscopic x-ray system, GE Healthcare, Salt Lake City, UT.
This recall has been terminated (originally issued September 28, 2007).
- Company
- Ge Oec Medical Systems, Inc
- Recall Initiated
- March 1, 2007
- Posted
- September 28, 2007
- Terminated
- May 1, 2012
- Recall Number
- Z-0676-2007
- Quantity
- 19 units
- Firm Location
- Salt Lake City, UT
- Official Source
- View on FDA website ↗
Reason for Recall
During routine service, primary and/or secondary collimators may not have been properly reinstalled.
Distribution
Nationwide and Germany.
Lot / Code Info
Serial Numbers: E2-0019, E2-0028, E2-0064, E2-0142, E2-0156, E2-0170, E2-0180, E2-0213, E9-0008, E9-0040-G, E9-0103, E9-0029, ES-0060, ES-0088, ES-0096, ES-0099, E9-0150, E9-0152, E9-0158.
Root Cause
Other
Action Taken
Customers were notified by letter on 03/01/2007. They were told to continue using the machines and that an inspection would be scheduled immediately to verify that collimators had been correctly installed.