X-Sizer Thrombectomy Catheter System (ev3) Model number XD-CS1150-45 (for US distribution). Model number XR-CS1150-45 (for foreign distribution). 1.5mm. Sterilization with Ethylene Oxide Gas. ev3 4600 Nathan Lane North, Plymouth, MN 55442-2920 USA.

This recall has been terminated (originally issued June 14, 2006).

Recall Initiated
February 14, 2006
Posted
June 14, 2006
Terminated
December 26, 2006
Recall Number
Z-1107-06
Quantity
Model number XD-CS1150-45 (US) 534 devices. Model number XR-CS1150-45 (OUS) 504 devices.
Firm Location
Plymouth, MN
Official Source
View on FDA website ↗

Reason for Recall

Loss of Operating Vacuum during use-It has been determined that a small percentage of X-Sizer Thrombectomy Catheter Systems from specific lots may lose operating vacuum prematurely due to a system leak within the Control Module. This failure can lead to the inaility to complete the procedure of thrombus removal and the necessity to remove the device prior to achieving a satisfactory result.

Distribution

World wide-AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MN, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. OUS to: Argentina, Australia, Austria, Brazil, Canada, Chile, Columbia, Czech Republic, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Jordan, Mexico, Netherlands, Norway, Pakistan, Portugal, Saudi Arabi, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom

Lot / Code Info

Model number XD-CS1150-45 (for US distribution), lot numbers: 782344, 782345, 782366, 782379, 782386, 782389,  782403, 782404, 782406, 782407, 782412, 782421, 782428,  782429, 782430, 782436, 782438, 782439, 782441, 782445,  782526, 986604, 986604, 986604, 986605, 986609, 986620,  986626, 986640, 986641, 986642, 986645, 986656, 986659,  1174394, 1174395, 1212529, 1229730, 1231296, 1234273,  1240986, 1241067, 1271238, 1272506, 1280664, 1280952,  1281088, 1282693, 1289590, 1289978, 1293902, 1295372,  1300605, 1302005, 1303572, 1308536, 1310238, 1362165.   Model number XR-CS1150-45 (for OUS distribution), lot numbers: 782773, 782777, 782787, 782788, 782790, 782791,  782795, 782796, 782797, 782798, 782800, 782801, 782804,  782808, 782811, 782812, 782816, 782818, 782822, 782848,  987054, 987055, 987056, 987058, 987059, 987062, 987063,  987064, 987065, 987066, 987067, 987070, 987070, 987070,  987071, 987073, 987076, 987078, 1057871, 1077841, 1077842, 1077843, 1077844, 1077846, 1077847, 1077848,  1077854, 1077856, 1077859, 1077862, 1095087, 1095089,  1296437, 1296552, 1313297, 1313502, 1313543, 1313562,  1313611, 1313645, 1314989, 1314991, 1314994, 1315002, 1315042, 1315058, 1315113, 1315140, 1315183, 1315211,  1315245, 1315292, 1315294, 1315313, 1315328, 1315365,  1315378, 1315383, 1315623, 1315624, 1315625, 1315626,  1315632, 1315633, 1315636, 1315639, 1315640, 1315644,  1315646, 1317128, 1317133, 1317134, 1317135, 1317144,  1317301, 1326969, 1326969, 1326969, 1326970, 1326972,  1326973, 1326974, 1326977, 1326982, 1326983, 1326998,  1326999, 1327006, 1327008, 1327011, 1327546, 1342794,  1342852, 1342874, 1342882, 1342888, 1342957, 1342980,  1343007, 1343019, 1343036, 1343045, 1343094, 1343108,  1343125, 1343136, 1343266, 1343273, 1343292, 1343314,  1363701, 1363712, 1363725, 1363742, 1363791, 1363827,  1363850, 1363974, 1363976, 1364006, 1364034, 1364049.

Root Cause

Other

Action Taken

A medical Device Recall Letter dated 02/13/06 was sent to hospitals along the affected device lot numbers the hospital should have received on a Device Reconciliation Form. ev3 is asking for affected devices to be returned. Returned devices will be replaced at no cost.